Search results for "Single molecule counting"
showing 1 items of 1 documents
Establishing the 99th percentile for high sensitivity cardiac troponin i in healthy blood donors from southern italy
2019
Introduction: The knowledge of high sensitivity cardiac troponin I (hsTnI) distribution in a reference population is mandatory for its introduction in clinical practice. The aim of this study was to define the Upper Reference Limit (URL) of hsTnI measured by Single Molecule Counting technology (SMC) in an accurately selected reference population. Materials and methods: In the study 1140 blood donors were included and selected on the basis of medical history and biomarkers. High sensitivity cardiac troponin I was measured by SMC technology (Clarity, Singulex, Alamed, USA). The 99th percentile was calculated by the non-parametric method according to the Clinical and Laboratory Standard Instit…